It is used to test the material properties of substances under high-temperature conditions, as well as the reaction rate, reaction kinetics, and activation energy between different reactants under high-temperature.
Adopts a new double-helix graphite heating furnace design, achieving a heating temperature of ≥1700℃. It features a long uniform temperature zone, long service life, and easy maintenance.
The furnace interior evacuation or inert gas filling is fully automated with one-key operation.
Test Atmospheres: Oxidizing, reducing, inert, and water vapor.
Evolved Gas Analysis: Equipped with dedicated constant-temperature connectors, connecting belts, and constant-temperature controllers for mass spectrometry (MS) and infrared (IR) analysis.
Data Acquisition and Analysis Software: Includes comprehensive data acquisition and analysis functions, text program editing, power-off data storage, 64 segments of temperature control programs, up to 99 experimental cycles, real-time data analysis, multi-curve comparison, storage and export of evaluation data, input and output of ASCII data, and data export to Microsoft Excel.
During system data collection, screenshots can be captured at any time, and the measurement range is automatically switched according to the output signal intensity.
Users can perform temperature and thermogravimetric (TG) calibration using standard samples.
Large-screen LCD display for real-time presentation of instrument status and data; equipped with two sets of thermocouples: one displays the furnace temperature in real time (regardless of whether the heating furnace is in operation), and the other displays the sample temperature during operation.
Temperature Parameters | |
Temperature Range | RT ~ 1700℃ (Capable of long-term constant temperature operation and equipped with a furnace self-protection system) |
Temperature Accuracy | ±0.1℃ |
Heating/Cooling Rate | 0.1℃/min~100℃/min |
Temperature Focus Area | 100mm |
Balance Parameters | |
Balance Type | Adopts a mechanical null photoelectric balance, featuring high precision and good repeatability |
Thermogravimetric Range | (0-10g adjustable; other ranges can also be customized) |
Balance Sensitivity | 0.1μg |
Thermogravimetric Noise | <0.1μg |
Crucible Configuration | Ceramic Crucible |